MICROPLATE READER Model DNM-9602 |
|
Main Function |
The instrument is compatible with registered diagnostic EIA kit, primarily used in hospitals or some detection related to clinical immunology.
|
Function and Technical Specifications |
|
Fundamental Parameters
- Range of working wavelength is from 400nm to 800nm ;
- Range of detection is from 0.000A to 4.000A ;
- Resolving power 0.001 A ;
- Filters when leaving factory 405nm, 450nm, 492nm, 630 nm, with the maximum array of 10 filters between 400nm - 800nm simultaneously.
- Reading speed: single wave :5s, dual waves:10s;
- Dimension: 420×325×210 mm (L×H×W)
- Weight: 9 kg;
- Monitor: 240*128 LCD
- Printer: Built in Printer
- Reader Speed :20s/96 Well
- Vibration Plate function :Strong/Weak: two gears
- Micro plate types :Should be able to read 6-,12-,24-,48-and 96-well micro plates
|
|
Technical Specifications:
1. Characters of the filters
Table 1: Characters of the Filters
Accuracy of wavelength (nm) |
Half-width (nm) |
±3.0 |
≤12 |
2.
The errors of absorbance accuracy are as follows
Table 2: Absorbance Accuracy
Absorbance (A) |
Accuracy Error |
0.000-2.000 |
±0.01 |
> 2.000-3.000 |
±0.05 |
> 3.000-4.000 |
±0.15 |
3. The Linearity Errors are as follows:
Table 3: Linearity of the Instrument
Absorbance (A) |
Error (%) |
0.000-2.000 |
±1.0% |
> 2.000-3.000 |
±2.0% |
> 3.000-4.000 |
±5.0% |
4. Reproducibility Error: ≤0.5%;
5. Stability Error: ≤ ± 0.005A/10min |
Main Features: |
- Provided with power on self-checking. Error data will display on LCD monitor if something wrong with the photoelectric system.
- Filter choosing-mode: automatically
- Detection mode: single wave/dual wave detection; single aperture /dual aperture detection
- Input mode: key in film keyboard;
- Input format: input capital letter/lowercase letter/number at random;
- Item parameters: Provided with function of set/save/ modify/ inquiry 126 calibration parameters and curves, to make detecting course more conveniently.
- Defection of calibration: with the function of calibration-defection, which can save/modify 126 parameters of calibration and curves.
- Detection methods: absorbance /qualitative /quantitative detection
- Saving function: can save the detection results on 105 pieces of plates
- Output mode:
|
|
Absorbance: |
display/printout originally absorbable value, |
|
Qualitative checking: |
display/printout absorbable value, critical value, S/CO, qualitative results; |
|
Quantitative checking: |
display/printout absorbable value, density, standard reference value, quantitative results |
- QC function: inquiry/printout the QC data and chart
- Communication: equipped with RS-232 interface.
- With self-monitoring function (real-time) on each testing light-source signal channel.
- With dormancy function on light-source lamp.
- With a CE Marked
|
|
Operating Conditions |
- Environmental temperature: 10 ~ 30℃;
- Relative humidity: ≤70%;
- Power Supply : 220 x (1±10%)V ; 50 ×(1±2%) Hz; Fuse φ5×20 2A; Input power: 100VA
- Keep away from strongly electromagnetic field and avoid direct irradiation by strong light;
- Dustproof and shockproof on worksite; no aggressive gas; pollution level II
- Power supply must be firmly grounded. Transient over voltage level II
- Preheating time: 10min
- Don’t use instrument in damp environment.
- Use CVT or Servo Controlled Voltage stabilizer for safe power & safety of equipment
|